799
Filling rate
Rate at which the dosing cylinder is to be filled after titration. The maxi-
mum filling rate depends on the cylinder volume of the exchange unit or
dosing device used.
Titrando
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
Titrino
Range
0.01 ... 150.00 mL/min
Selection
maximum
Default value
maximum
5.6.3.5.2.7
KFT Ipol - Conditioning
Tab: Method
▶ KFT Ipol ▶ Properties... ▶ Conditioning
Command name
Name of the command.
Input
25 characters
Conditioning
Conditioning
on | off (Default value: off)
Switching conditioning on/off. If conditioning is deactivated all following
parameters are not visible.
Start drift
The determination can only be started when the currently measured drift
is less than the start drift entered here. In such cases, the Status in the
Run subwindow displayed on the Single determination or Determina-
tion series tab will then switch to COND READY.
Titrando
Range
1 ... 999 µg/min
Default value
20 µg/min
Drift correction
If this parameter is set to automatic or manual the drift correction ( =
Drift value × Time for drift correction) caused by empty consumption
during titration is deducted from the total consumption. The Time for