3 properties - tower, Properties - tower # 1 – Metrohm tiamo 2.3 Patch 1 User Manual
Page 1557
■■■■■■■■■■■■■■■■■■■■■■
7 Instruments
tiamo 2.3 Patch 1
■■■■■■■■
1541
Input
1000 characters
Default value
empty
7.13.3
Properties - Tower #
Tab: Configuration
▶ Devices ▶ [Edit] ▶ Properties... ▶ Properties - 'Device type' -
'Device name'
▶ Tower #
On the tab Tower the properties of the tower and of the robotic arm
connected to are displayed.
Tower parameters
Max. stroke path
Entry of the lowest permitted lift position. A lift height of 0 mm corre-
sponds to the "Home position", i.e. the lift is moved right to the top.
Range
0 ... 235 mm
Default value
235 mm
Min. beaker radius
Definition of the minimum radius which the beakers used on the rack
must have. If the lift is moved to the work position the beaker radii
defined in Rack table(see Chapter 6.12.2.1, page 1454) for general
Sample positions(see Chapter 6.12.3.2, page 1457) and Special beak-
ers(see Chapter 6.12.3.4, page 1459) will be compared with the Min.
beaker radius. If this minimum beaker radius is undercut then the run
will be stopped and an error message will appear. With off no check will
be made.
Selection
1.0 … 100.0 mm | off
Default value
off
Lift rate
Lift speed for manual operation.
Range
3 ... 25 mm/s
Default value
25 mm/s
Axial distance
Distance between the axis of rotation of the sample rack and swing axis of
the robotic arm.
Range
100.0 ... 300.0 mm
814 USB Sample Processor XL
Selection
166.0 mm
Default value
166.0 mm