3 tet - start conditions, Start conditions – Metrohm tiamo 2.3 Patch 1 User Manual
Page 934
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
918
■■■■■■■■
tiamo 2.3 Patch 1
not defined
If the sensor is not defined, then it will be checked at the start of the
determination whether a sensor of the type Thermoprobe is connec-
ted to the measuring input. If this is the case, the connected sensor is
used; if it is no the case, an error message appears.
Stirrer
Stirrer
Selection of the stirrer.
Selection
1 | 2 | 3 | 4 | off
Default value
1
off
No stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Range
-15 ... 15
Default value
8
Switch on/off automatically
on | off (Default value: on)
If this option is activated, the stirrer will be switched on automatically at
the start of the command and will be switched off automatically at the
end of the command.
5.6.3.9.3
TET - Start conditions
Tab: Method
▶ TET ▶ Properties... ▶ Start conditions
Command name
Name of the command.
Input
25 characters
The start conditions are processed in the listed sequence before the titra-
tion is started.
Start volume
Start volume
Volume to be dosed with the indicated dosing rate before the start of the
titration.
Range
0.00000 ... 50.0000 mL
Default value
0 mL