5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
916
■■■■■■■■
tiamo 2.3 Patch 1
Designa-
tion
Description
MI.TIM
Time at minimum measured value in s
MI.VOL
Volume at minimum measured value in mL
NMP
Number of measuring points in measuring point
list
SME
Start measured value (measured value after pro-
cessing the start conditions) in the unit of the
measured value
STY
Type of stop with which the command was stop-
ped: 1 = normal; 0 = manual or after error
SVA
Start volume absolute (volume that was added
according to the start condition "start volume")
in mL
TITER
Titer value of the solution used for the command
5.6.3.9.2
TET - General/Hardware
Tab: Method
▶ TET ▶ Properties... ▶ General/Hardware
Command name
Name of the command.
Input
25 characters
The general parameters for the device, the electrodes and the stirrer are
defined on this tab.
Device
Device name
Selection of a device from those available in the device table. The only
devices which shall be offered are those with which the command can
run.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.
Device type
Display or selection of the device type. If a device is selected under Device
name, then this Device type field can no longer be edited, and the
device type belonging to the selected device is displayed instead.