4 curve properties - y2 axis, Y2 axis – Metrohm tiamo 2.3 Patch 1 User Manual
Page 408
4.8 Subwindow Curves
■■■■■■■■■■■■■■■■■■■■■■
392
■■■■■■■■
tiamo 2.3 Patch 1
4.8.4.4
Curve properties - y2 axis
Tab: Database
▶ Curves ▶ Properties curves # ▶ y2 axis
Parameters for the graphical display of the curves on the y2 axis (right-
hand y axis).
Command type
Selection of the command type for which the curve properties are to be
defined. The curve properties defined for each command type are saved
per curve window and per client.
Selection
DET | MET | SET | MEAS | MEAS T/Flow | MEAS
TC Cond | MEAS Ref | MEAS TMF | MEAS Spec
| MEAS Opt | CAL Spec | KFT | KFC | BRC | STAT
| DOS | TET
Default value
DET
Note
When the dialog window is opened, those Command type will be
selected per default which applies for the curve shown in the curve
window.
Autoscaling
on | off (Default value: on)
If this option is switched on then all axes in the curve window will be
scaled automatically. In this case the fields Start value and End value
cannot be edited.
Quantity
Selection of the quantity to be shown on the y2 axis.
For DET commands
Selection
Volume [mL] | Measured value | ERC | Time [s] |
Temperature [°C] | Calculated 1...3 | External
1...3 | off
Default value
off
For MET commands
Selection
Volume [mL] | Measured value | dMW | Time
[s] | Temperature [°C] | Calculated 1...3 | Exter-
nal 1...3 | off
Default value
off