4 met cond - titration parameters, Titration parameters – Metrohm tiamo 2.3 Patch 1 User Manual
Page 697
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1
■■■■■■■■
681
Start volume
Start volume
Volume to be dosed with the indicated dosing rate before the start of the
titration.
Range
0.00000 ... 9999.99 mL
Default value
0.00000 mL
Dosing rate
Rate (volume/time unit) at which the start volume is to be added. The
maximum dosing rate depends on the cylinder volume of the exchange
unit or dosing unit used.
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
Pause
Pause
Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the addition of the start volume. The pause follows at
the end of all the start conditions.
Range
0 ... 999999 s
Default value
0 s
5.6.3.3.6.4
MET Cond - Titration parameters
Tab: Method
▶ MET Cond ▶ Properties… ▶ Titration parameters
Command name
Name of the command.
Input
25 characters
Parameters defining the run of the titration.
Titration rate
Titration rate
The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for the Measured value
acceptance and Dosing of increments of these sets are not displayed.
The User setting must be selected in order to be able to edit these param-
eters.
Selection
slow | optimal | fast | User
Default value
optimal