Metrohm tiamo 2.3 Patch 1 User Manual
Page 1059
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1
■■■■■■■■
1043
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.
Device type
Display or selection of the device type. If a device is selected under Device
name, then this Device type field can no longer be edited, and the
device type belonging to the selected device is displayed instead.
If the option not defined is selected as Device name, any device types
or device groups which are able to execute the command can be selected,
independently of the devices in the device table.
Selection
089 Photometer | Avantes Spectrometer
Default value
Avantes Spectrometer
Sensor
The section Sensor is only displayed for Device type = 089 Photometer.
Temperature measurement
Type of temperature measurement
Selection
continuous | off
Default value
continuous
continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Measuring parameters tab will be
used.
Stirrer
The section Stirrer is only displayed for Device type = 089 Photometer.
Stirrer
Selection of the stirrer.
Selection
1 | off
Default value
1
off
Means that no stirrer will be used.