5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
446
■■■■■■■■
tiamo 2.3 Patch 1
Conditioning
Automatic conditioning
on | off (Default value: on)
If this option is enabled, the conditioning will be started automatically at
the start of a determination for all commands with activated conditioning
(SET, KFT, KFC and BRC). After the track with the conditioning com-
mand has been finished, conditioning is started again automatically and
will be continued even after the end of the determination.
If this parameter is disabled, then conditioning will not be started until
the start of the command. Conditioning is not started again after the end
of the track with the conditioning command. This parameter must be
switched off for automation applications.
5.6.2.2.3
START - Application note
Tab: Method
▶ START ▶ Properties... ▶ START 'Command name' ▶ Application
note
Command name
Name of the command.
Input
25 characters
The application note defined in the text window is displayed in the tab
Application note in the subwindow Live display of the program part
Workplace. The tab is opened automatically when the method is loaded.
The text editor with which the application note can be entered or
changed is opened with
or by double-clicking into the text field (see
5.6.2.2.4
START - Method variables (Table)
Tab: Method
▶ START ▶ Properties... ▶ START 'Command name' ▶ Method varia-
ble
Command name
Name of the command.
Input
25 characters
The variables that are to be available within the method in other com-
mands (e.g. CALC command) under the designation MV.'Variable
name' must be declared here. Either a sample data variable (entry in field
Assignment) or a direct value (entry in field Value) can be assigned to a
method variable.