6 stat ph - stop conditions, Stop conditions – Metrohm tiamo 2.3 Patch 1 User Manual
Page 892
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
876
■■■■■■■■
tiamo 2.3 Patch 1
Start time
The measuring points are not entered in the measuring point list until this
time has elapsed since the start of the titration.
Range
0 ... 999999 s
Default value
0 s
Start meas. value pH
The measured values are not entered in the measuring point list until this
value has been reached.
Titrando, 855
Range
-20.000 ... 20.000
Selection
off
Default value
off
Titrino
Range
-20.00 ... 20.00
Selection
off
Default value
off
Start rate
The measured values are not entered in the measuring point list until the
dosing rate has fallen below this value. This parameter is not checked until
10 s after the start.
Titrando, 855
Range
0.01 ... 166.00 mL/min
Selection
off
Default value
off
Titrino
Range
0.01 ... 150.00 mL/min
Selection
off
Default value
off
5.6.3.8.6.6
STAT pH - Stop conditions
Tab: Method
▶ STAT pH ▶ Properties... ▶ Stop conditions
Command name
Name of the command.
Input
25 characters
Conditions for stopping the titration. If more than one stop condition is in
effect at the same time, then the criterion which is fulfilled first will stop
the titration.