5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1150
■■■■■■■■
tiamo 2.3 Patch 1
Stirrer
Stirrer
Selection of the stirrer.
Titrando, 855, 867
Selection
1 | 2 | 3 | 4 | off
Default value
1
off
Means that no stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Titrando, 855, 867
Range
-15 ... 15
Default value
8
Switch off automatically
on | off (Default value: on)
If this option is activated, the stirrer will be switched off automatically
when the command has finished. This parameter is displayed only for
Titrandos, 855 and 867.
5.6.5.13.3
CAL MEAS pH - Measuring parameters
Tab: Method
▶ CAL MEAS pH ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Input
25 characters
The measuring parameters for the calibration are defined on this tab.
Measurement with drift control
The pH calibration is carried out drift-controlled. The measurement is stop-
ped as soon as the Signal drift is reached or a defined waiting time has
elapsed.
Signal drift
The measured value is only accepted if the signal drift defined here has
been fallen short of.