5 Method
tiamo 2.3 Patch 1
■■■■■■■■
725
Min. waiting time
The minimum waiting period is only important for drift-controlled meas-
urements. The measured value is not accepted until the minimum waiting
time has elapsed, even if the signal drift has already been reached. The
drift continues to be checked while the waiting time is elapsing.
Range
0 ... 999999 s
Default value
0 s
Max. waiting time
If signal drift has been switched off or has not yet been reached, then the
measured value will be accepted when the maximum waiting time has
elapsed.
Range
0 ... 999999 s
Default value
1 s
Pause 1
Pause 1
Waiting time, e.g. until the electrode has settled down, before a start vol-
ume is added.
Titrando, 719, 720, 736, 751, 758, 785, 794, 798, 799
Range
0 ... 999999 s
Default value
0 s
Start volume
Start volume
Volume to be added before the start of the titration at the dosing rate
indicated.
Titrando, 855
Range
0.00000 ... 9999.99 mL
Default value
0.00000 mL
Titrino
Range
0.00 ... 999.99 mL
Default value
0.00 mL
Dosing rate
Speed at which the start volume is to be added. The maximum dosing rate
depends on the cylinder volume of the exchange unit or dosing unit used.
Titrando, 855
Range
0.01 ... 166.00 mL/min