10 met u - volume window – Metrohm tiamo 2.3 Patch 1 User Manual
Page 634
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
618
■■■■■■■■
tiamo 2.3 Patch 1
Titrando, 855
Range
-2000.0 ... 2000.0 mV
Default value
2000.0 mV
Titrino
Range
-2000 ... 2000 mV
Default value
2000 mV
EP criterion
Criterion for the recognition of endpoints. Endpoints whose discovered
ERC value is smaller than the value entered here will not be recognized.
Range
1 ... 999 mV
Default value
30 mV
EP recognition
Filter for the recognition of equivalence points:
Titrando, 855
Selection
first | greatest | last | ascending | descending
Default value
first
first
Only the first endpoint to be found will be recognized.
greatest
Only the endpoint with the greatest ERC value, i.e. the steepest jump,
will be recognized.
last
Only the last endpoint to be found will be recognized.
ascending
Only endpoints with a positive slope of the titration curve will be rec-
ognized.
descending
Only endpoints with a negative slope of the titration curve will be rec-
ognized.
5.6.3.3.3.10
MET U - Volume window
Dialog window: Method
▶ MET U ▶ Properties... ▶ Potentiometric evalua-
tion
▶ Evaluation with measured value window ▶ [New]/[Properties] ▶ Volume
window #
Volume windows are defined sub-regions (windows) on the volume axis
for which different parameters for the potentiometric evaluation can be
defined. Only those endpoints which lie within these windows and addi-
tionally meet the parameters defined for each window will be recognized.
Only one endpoint will be recognized per window.