9 met cond - volume window, 10 met cond - fixed endpoint evaluation – Metrohm tiamo 2.3 Patch 1 User Manual
Page 708
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
692
■■■■■■■■
tiamo 2.3 Patch 1
5.6.3.3.6.9
MET Cond - Volume window
Dialog window: Method
▶ MET Cond ▶ Properties… ▶ Conductometric evalua-
tion
▶ Evaluation with volume window (mL) ▶ [New]/[Properties] ▶ Volume win-
dow #
Volume windows are defined sub-regions (windows) on the volume axis
for which different parameters for the potentiometric evaluation can be
defined. Only those endpoints which lie within these windows and addi-
tionally meet the parameters defined for each window will be recognized.
Only one endpoint will be recognized per window.
Lower limit
Lower limit for the volume window.
Range
0.00000 ... 9999.99 mL
Default value
0.00000 mL
Upper limit
Upper limit for the volume window.
Range
0.00000 ... 9999.99 mL
Default value
9999.99 mL
Smoothing
Measure for the curve smoothing used for the data processing. The higher
the smoothing that is selected, the more the measurement curves will be
smoothed.
Range
0 ... 150
Default value
0
5.6.3.3.6.10
MET Cond - Fixed endpoint evaluation
Dialog window: Method
▶ MET Cond ▶ Properties… ▶ Additional evalua-
tions
▶ Fixed endpoint evaluation ▶ [New]/[Properties] ▶ Fixed endpoint evalua-
tion #
Quantity
Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.
Selection
Measured value | Time | Volume
Default value
Measured value
Fixed value
Value of the fixed endpoint.