6 Configuration
tiamo 2.3 Patch 1
■■■■■■■■
1407
If the warning or intervention limits are exceeded then the line number
will also be shown with an orange and red background respectively.
Calibration date
Date and time of the calibration.
Calibration method
Name of the method with which the calibration was carried out. If the cal-
ibration data has been entered manually, then manual will appear here.
Note
This parameter is not saved on the data chip of intelligent sensors, i.e.
this parameter will be empty if the data are written into the sensor
table from the sensor.
User
Short name of the user logged in during calibration or when the calibra-
tion data was entered manually. If work is not carried out using login then
the user logged in under Windows will be entered automatically.
Note
This parameter is not saved on the data chip of intelligent sensors, i.e.
this parameter will be empty if the data are written into the sensor
table from the sensor.
History graph
On the two tabs Slope and pH(0) or E(0) the last 10 values are shown
for the selected sensor. As in the history table, the values are shown in dif-
ferent colors:
■
Blue, if the value has been assigned automatically by a method.
■
Black, if the values have been entered manually.
If limits have been defined, then the warning limits will be shown in
orange and the intervention limits in red.
Note
The history graph can be copied into the clipboard by using the con-
text-sensitive menu item Copy.