Metrohm tiamo 2.3 Patch 1 User Manual
Page 679
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1
■■■■■■■■
663
Potentiometric evaluation
Selection
Evaluation without window | Evaluation with
measured value window (Upol) | Evaluation
with volume window (mL)
Default value
Evaluation without window
Evaluation without window
With this option set, the two parameters EP criterion and EP recog-
nition will be applied across the entire range of the titration curve.
Evaluation with measured value window (Upol)
With this option set, up to 9 regions (windows) can be defined on the
measured value axis. Only those equivalence points which lie within
these windows and additionally meet the parameters defined for each
window will be recognized. Per window only one EP will be recog-
nized. The defined windows with their parameters are shown in the
window table and can be edited with the following buttons:
Evaluation with volume window (mL)
This option is visible only for Titrando and 855.. With this option up
to 9 regions (windows) can be defined on the volume axis. Only those
equivalence points which lie within these windows and additionally
meet the parameters defined for each window will be recognized. Only
one EP will be recognized per window. The defined windows with their
parameters are shown in the window table and can be edited with the
following buttons:
Evaluation without window
EP criterion
Criterion for the recognition of equivalence points which is valid for all
windows. Equivalence points whose ERC is smaller than the set EP crite-
rion will not be recognized.
Range
0.1 ... 99.9 µA
Default value
2.0 µA
EP recognition
Filter for the recognition of equivalence points:
Titrino
Selection
all | greatest | last | off
Default value
all
Titrando, 855
Selection
all | greatest | last | ascending | descending |
off
Default value
all