5 Method
tiamo 2.3 Patch 1
■■■■■■■■
917
If the option not defined is selected as Device name, any device types
or device groups which are able to execute the command can be selected,
independently of the devices in the device table.
Selection
Device types | 859 Titrotherm
Default value
859 Titrotherm
Dosing device
Dosing device
Selection of the number of the dosing device (dosing unit) with which the
solution is to be dosed. All the dosing device connections which are possi-
ble with the selected device type are displayed.
Selection
1 | 2 | 3 | 4
Default value
1
Solution
Input of a solution name or selection of a solution from the solutions
defined in the solution table. If intelligent dosing units are used, then a
check will be made in the method run to see whether the correct solution
has been attached to the selected dosing device. With non-intelligent dos-
ing units, only the cylinder volume is checked. At the start of the com-
mand, a check is made of the working life, the validity of the titer and the
GLP test interval for the selected solution.
Selection
'Solution name' | not defined
Default value
not defined
not defined
No tests will be carried out.
Sensor
Measuring input
Selection of the measuring input the sensor is connected to.
Selection
1 | 2
Default value
1
Sensor
Selection of a sensor of type Thermoprobe from the sensors configured
in the sensor table. The calibration data for the sensor will be adopted for
the determination.
Selection
'Sensor name' | not defined
Default value
not defined