11 met ph - fixed endpoint evaluation, 3 met u, 1 met u - overview – Metrohm tiamo 2.3 Patch 1 User Manual
Page 610
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
594
■■■■■■■■
tiamo 2.3 Patch 1
5.6.3.3.2.11
MET pH - Fixed endpoint evaluation
Dialog window: Method
▶ MET pH ▶ Properties... ▶ Additional evalua-
tions
▶ Fixed endpoint evaluation ▶ [New]/[Properties] ▶ Fixed endpoint evalua-
tion #
Quantity
Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.
Selection
Measured value | Time | Volume
Default value
Measured value
Fix value
Value of the fix endpoint.
Measured value
Range
-20.000 ... 20.000 pH
Time
Range
0.0 ... 999999.9 s
Volume
Range
0.00000 ... 9999.99 mL
5.6.3.3.3
MET U
5.6.3.3.3.1
MET U - Overview
Dialog window: Dialog window
▶ MET U ▶ Properties... ▶ MET U - 'Command
name'
Command for Monotonic equivalence point titrations with potentio-
metric voltage measurement.
Instruments
This command can be executed with the following devices:
Titrando: 808, 809, 835, 836, 857, 888, 904, 905, 906, 907
Titrino: 702, 716, 721, 736, 751, 785, 794, 798, 799
Robotic Titrosampler: 855
Appearance
The command has the following appearance: