If the output is to be an impulse with a different length then an appropriate
template must be defined for it.
The output lines and bits are numbered from right to left:
Output 13 12 11 10 9 8 7 6 5 4 3 2 1 0
Bit
13 12 11 10 9 8 7 6 5 4 3 2 1 0
Examples:
************1*
sets output line 1 to active (= set) which would corre-
spond to a Stop command, e.g. for a connected Titrino.
************0*
sets the line to inactive.
Note
We recommend that non-relevant output lines are masked with an asterisk
*
so
that the status of these lines is not altered.
Note
For Titrinos with 3 output lines only 3 characters can be entered. If a signal tem-
plate is selected, only the first 3 characters are used.
Set the binary pattern defined under
Outputs
.
2.6.8
Sample changer functions
If in the subwindow for Select device a sample changer tower is selected then the
subwindow Functions/Parameters opens and shows the functions that are possi-
ble with the sample changer together with their associated parameters.
Sample changer functions
The sample changer functions are shown on the following tabs:
•
General
•
Move
•
Assign position
•
Pump
•
Heater/Gas
Devices
The sample changer functions can be carried out with the following instruments:
730, 774, 778, 789, 814, 815, 855