Subwindow Determination overview
243
guaranteed that the method will proceed in the same way during reproc-
essing as during the determination itself.
•
Monitoring
If the method contains monitoring of variables and results then any mes-
sages which may occur will be shown and included in the report, but no e-
mails will be sent nor will beeps be produced. If one of the two actions
Stop
determination
or
Stop determination and series
is switched on then reproc-
essing will be canceled.
•
Titration and measuring commands
For DET, MET, SET, KFT, KFC, MEAS, CAL MEAS only the evaluations will
be recalculated with the existing data, all other parameters will be ignored.
For commands that generate data, which are carried out during the deter-
mination but are no longer carried out during reprocessing, all the existing
values will be deleted. If commands that generate data are carried out for
the first time during the recalculation then the recalculation will be termi-
nated.
•
Waiting and timeout periods
All waiting and timeout periods in commands will be set to the value
0
for
the recalculation.
•
Non-executed commands
The commands REPORT, REQUEST, WAIT, RECEIVE, SEND, TRANSFER,
STIR, SCAN, CTRL, MOVE, SWING, LIFT, PUMP, RACK, HEATER, FLOW,
ADD, LQH, PREP and EMPTY will be performed but not executed.
•
Command variables
If a method contains calculations using command variables the determina-
tion can only be recalculated correctly if the DATABASE command is exe-
cuted after the corresponding commands.
•
Solution variables
If titrants are altered in the method then their associated variables (e.g.
Titer
,
Slope
) will always (even when at the same time the corresponding
variables have been altered under
Variables
) be taken from the Configura-
tion data and updated for the recalculation.
•
Assign titer and common variables
If a determination without statistics is recalculated in which a titer or a
common variable has been assigned, then the assignment will only take
place when the recalculation is confirmed with
[OK]
. If a determination with
statistics is recalculated in which a titer or a common variable has been as-
signed, then the assignment with the data of the last recalculated determi-
nation will only take place after the recalculation is confirmed with
[OK]
.
Modifications
In the subwindow
Modifications
of the dialog window
Reprocessing
modifications
can be made to the following 4 tabs:
•
Variables
Modification of the variables used in the selected determination.
•
Method
Modification of the method used in the selected determination.
•
Statistics
Modification of the statistics data.
•
Curve evaluation
Manual modification of the curve evaluation.