Metrohm tiamo 1.3 Manual User Manual
Page 582

556
Method
Manually entered measuring temperature. If a temperature sensor is con-
nected and
temperature measurement
on the tab
General/Hardware
is set to
automatic
or
continuous
, the temperature is measured continuously. The
value is used for the temperature adjustment during the measurement.
Time interval measuring point
0.1 ... [ 5.0 ] ... 999'999 s
(Titrando)
1 ... [ 5 ] ... 999'999 s
(Titrino)
Time interval for entering a measuring point in the list of measuring points.
Stop conditions
Conditions for stopping the dosing. The stop conditions are checked at the same
time and the condition which is fulfilled first stops the titration.
Command name
25 characters, [ DOS pH # ]
Name of the command.
Stop volume
0.00000 ... [ 100.000 ] ... 9999.99 mL, off
(Titrando)
0.000 ... [ 100.000 ] ... 9999.99 mL, off
(Titrino)
Stops when the given volume has been added after the start of the dosing.
The stop volume should be adapted to suit the sample weight or the titra-
tion vessel size.
Stop time
0 ... 999999 s, [ off ]
(only Titrando)
Stops when the preset time has elapsed since the start of the dosing.
Filling rate
0.01 ... 166.00 mL/min, [ maximum ]
(Titrando)
0.01 ... 150.00 mL/min, [ maximum ]
(Titrino)
Speed at which the dosing cylinder is to be filled after dosing. The maxi-
mum filling rate depends on the cylinder volume of the exchange unit or
dosing unit used. In case the entered filling rate is too high for the selected
dosing device, the rate will be reduced automatically to the highest possi-
ble value.
Note
Lower filling rates should be entered for solutions with higher viscosities.