Metrohm tiamo 1.3 Manual User Manual
Page 545
Commands
519
Evaluations
The following methods for evaluating measuring curves are defined on this tab.
Command name
25 characters, [ MEAS Cond # ]
Name of the command.
Fix end point evaluation
on, [ off ]
To a fixed quantity (measured value or time) the associated values for the
other quantities are interpolated from the list of measuring points for the
fixed end point.
Fixed quantity
[ Measured value ], Time
Selection of the fixed quantity to which the associated values for the other
quantities are to be interpolated from the list of measuring points.
Fix EP# at
0.0000 ... 2'000.0 mS/cm, [ off ]
Fixed end point 1 ... 9 for
Fixed quantity
=
Measured value
.
Fix EP# at
0.0 ... 999'999.9 s, [ off ]
Fixed end point 1 ... 9 for
Fixed quantity
=
Time
.
The fixed end points Fix EP4...9 are displayed additionally to Fix EP1...3 by
pressing this button.
Only the fixed end points Fix EP1...3 are displayed by pressing this button.
Minimum evaluation
on, [ off ]
For the minimum measured value the associated time and temperature are
interpolated from the list of measuring points) (see Titration commands -
Evaluation - Minimum/Maximum evaluation).