526

Method

Command name

25 characters, [ STDADD man # ]

Name of the command.

Measurement with drift control

The measurement is stopped as soon as the

Signal drift

or the

Stop meas-

ured value

is reached or a

waiting time

has elapsed.

Signal drift

0.1 ... [ 10.0 ] ... 999.0 mV/min

The measured value is only accepted if the signal drift defined here has
been fallen short of.

Min. waiting time

[ 0 ] ... 999'999 s

The measured value is only accepted if the minimum waiting time has

elapsed even if the signal drift has already been reached. The drift contin-
ues to be checked as the minimum waiting time is passing.

Max. waiting time

0 ... [ 52 ] ... 999'999 s

If the signal drift has not yet been reached then the measured value will be
accepted when the maximum waiting time has elapsed. If no new entry for

the waiting time has been made then a suitable value for the drift will be
calculated automatically according to the following equation:

Measuring interval

0.1 ... [ 2.0 ] ... 999'999.0 s

Time interval for entering a measuring point in the list of measuring points.