Metrohm tiamo 1.3 Manual User Manual

Page 558

background image

532

Method

opened with the menu item

Determinations, Show calibration curve...

or the symbol

.

General/Hardware

Screenshot and parameters: see STDADD dos - General/Hardware

Standard addition

The standard addition parameters are defined on this tab.

Command name

25 characters, [ STDADD auto # ]

Name of the command.

Number of additions

1 ... [ 3 ] ... 19

Number of times the standard solution is to be added.

Volume auxiliary solution

[ 0.000 ] ... 9999.999 mL

Volume of auxiliary solution (e.g. buffer) that is to be added to the sample

before the first standard addition.

Stop volume

0.0 ... [ 100.0 ] ... 9999.9 mL, off

Stops when the sum of the added volume increments exceeds the stop
volume defined here.

Addition solution

(only displayed for

Solution

=

not defined

)

Concentration

0.001 ...[ 1.000 ] ... 999999.999

Concentration of addition solution.

Unit