Metrohm tiamo 1.3 Manual User Manual
Page 801

Subwindow Titrants/Solutions
775
Titer history
History table
The table contains the last 10 titer determinations for the selected solution and
can neither be edited nor sorted. The titer determinations are arranged chrono-
logically so that the most recent determination is shown last.
Titer
Titer values are shown in the following colors:
- Blue, if the titer has been determined automatically by a method.
- Black, if the titer has been entered manually.
- Orange, if the titer is outside the warning limits.
- Red, if the titer is outside the intervention limits.
If the warning or intervention limits are infringed then the line number will
also be shown with an orange and red background respectively.
Date titer det.
Date and time of titer determination.
Titer method
Name of the method with which the titer determination was carried out. If
the titer was entered manually then
manual
will be shown here.
Note
This parameter is not saved on the datachip of intelligent exchange/dosing units,
i.e. this parameter will be empty if the data are written into the solutions table
from the IEU/IDU.
User
Short name of the user logged in during titer determination or when the titer
was entered manually. If work is not carried out using login then the user
logged in under Windows will be entered automatically.