beautypg.com

Metrohm tiamo 1.3 Manual User Manual

Page 557

background image

Commands

531

[ 0.00000 ] ... [ 0.100 ] ... 99999.9 mL

Quantity of the defined volume increments.

Addition solution

(only displayed for

Solution

=

not defined

)

Concentration

0.001 ...[ 1.000 ] ... 999999.999

Concentration of addition solution.

Unit

mol/L, mmol/L, µmol/L, [ ppm ], g/L, mg/L, µg/L, mg/mL, µg/mL, %, mEq/L

Concentration unit of addition solution.

Dosing rate

0.01 ... 166.00 mL/min, [ maximum ]

Speed at which the volume increments are added. The maximum dosing
rate depends on the cylinder volume of the used exchange unit or dosing
unit.

Measuring parameters

Screenshot and parameters: see STDADD man - Measuring parameters

STDADD auto

Command for Standard addition with automatic addition of the standard addi-

tion solution from a dosing device in such a way that a constant potential differ-
ence results.

Devices

This command can be executed with the following devices:
Titrando: 808, 809, 835, 836, 855, 857

Appearance

The command has the following appearance:

Parameters

The parameters for the command

STDADD auto

are configured on the following 3

tabs:

General/Hardware

Parameters for device, dosing device, sensor and stirrer..

Standard addition

Parameters for standard addition.

Measuring parameters

Parameters for the measurement.

Standard addition curve

The standard addition curve with its evaluation data for a determination is shown

in the program part Database in the dialog window

Calibration curve

which is