572

Method

Volume

0.00000 ... [ 10.00000 ] ... 99'999.9 mL (Dosing)
0.00000 ... [ 10.00000 ] ... 50.0 mL (Aspirate)

Enter the volume to be processed. This parameter is only visible for the

Functions

Dosing

and

Aspirate

.

Rate

[ maximum ]
0.01 ... 166.00 mL/min

Speed at which the commands

Dosing

,

Fill

,

Aspirate

,

Eject

and

Exchange po-

sition

are carried out. The maximum rate depends on the cylinder volume of

the Dosing unit used. Aspirating and ejecting the sample should be done
with a rate < 10 mL/min.

PREP

Command for rinsing cylinder and tubing of an exchange or dosing unit.

Devices

This command can be executed with the following devices:

Titrino: 736, 751, 758, 799

Titrando: 808, 809, 835, 836, 841, 842, 855, 857, 888, 890

Dosing Interface: 846

Sample Processor: 730, 774, 778, 789, 814, 815, 855