Metrohm tiamo 1.3 Manual User Manual
Page 470

444
Method
•
Keep filling times short
Use the highest possible filling rate in order to keep the filling times as short
as possible. Please note the viscosity and density of the liquid.
•
Filling rate with different cylinders
When you use two dosing devices with different cylinder volumes (Dosing
cylinder 2 > Dosing cylinder 1) the filling rate of the larger cylinder must be
at the minimum:
Filling rate 2 >= Filling rate 1
•
(V
Dosing cylinder 2
/ V
Dosing cylinder 1
)
•
Rule for dosing rate
The dosing rate must not exceed 75 % of the filling rate of the smaller cylin-
der. This corresponds to the following values (at the maximum filling rate):
Cylinder volume
Max. dosing rate
Max. flow rate
Exchange unit
Dosing unit
Exchange unit
Dosing unit
1 mL
2.25 mL/min
---
ca. 130 mL/h
---
2 mL
---
5.00 mL/min
---
ca. 300 mL/h
5 mL
11.25 mL/min
12.50 mL/min
ca. 670 mL/h
ca. 750 mL/h
10 mL
22.50 mL/min
25.00 mL/min
ca. 1.3 L/h
ca. 1.5 L/h
20 mL
45.00 mL/min
50.00 mL/min
ca. 2.7 L/h
ca. 3.0 L/h
50 mL
112.50 mL/min
124.50 mL/min
ca. 6.7 L/h
ca. 7.5 L/h
Sensor
Measuring input
[ 1 ] ... 2
(835, 836, 842, 857)
[ 1 ]
(855)
[ 1 ] ... 2, diff.
(Titrino)
Selection of the measuring input to which the sensor is connected.
Sensor
Sensor name, [ pH electrode ]
Selection of a sensor of type
pH electrode
from the list of sensors defined in
the sensor table. The calibration data for the sensor will be adopted for the
determination.
Temperature measurement
Type of temperature measuring (only Titrando):
continuous
A temperature sensor must be connected. The temperature is measured
continuously.
[ automatic ]
If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise the manually entered
Temperature
(tab
Titra-
tion parameters
) will used.
off
The temperature will not be measured. The
Temperature
(see Titration pa-
rameters) which has been entered manually will be used.
Stirrer
Stirrer
[ 1 ] ... 4, off
(only Titrando)
Selection of the stirrer.
off
means that no stirrer will be used.
Stirring rate
-15 ... [ 8 ] ... 15
(only Titrando)
Setting the stirrer speed. The direction in which stirring takes place changes
as the sign in front of the stirrer speed changes.
Switch off automatically