Metrohm tiamo 1.3 Manual User Manual
Page 429
Commands
403
Sample addition time
[ 0 ] ... 99 s
Waiting time between stop of conditioning and start of titration during which
the sample can be added.
Only start titration by a start command from a SEND command
on, [ off ]
If this option is set the titration is only started when the
Event message
Start
titration
from a
SEND
command reaches this command.
Note
At the time of reception of the
SEND
command conditioning must already be
terminated (=
Condok
) otherwise the command to start the titration will be ig-
nored.
Show measured value during conditioning
on, [ off ]
If this option is set additionally to the drift and time the measured value is
shown in the Live display.
Additional evaluations
The following additional methods for evaluation of titration curves can be activated
and defined here:
Command name
25 characters, [ SET pH # ]
Name of the command.
Fix end point evaluation
on, [ off ]
For a fixed quantity (measured value, volume or time) for the fix end point
the associated values are interpolated from the list of measuring points.
Fixed quantity
[ Measured value ], Time, Volume
Selection of the fixed quantity to which the associated values for the other
quantities are to be interpolated from the list of measured points.
Fix EP# at pH
-20.000 ... 20.000, [ off ]
Fix end point 1 ... 9 for
Fixed quantity
=
Measured value
.