Metrohm tiamo 1.3 Manual User Manual
Page 589
Commands
563
Device
Device name
Device name, [ not defined ]
Selection of a device from the device table. Only those devices being able
to execute the command are selectable. Additionally,
not defined
can be se-
lected. In this case, the device name must be assigned by the user at the
start of the method.
Device type
Device types, [ Titrando ]
Display or selection of the device type. If a device is selected under
Device
name
, this parameter is not editable any more. The device type belonging to
the selected device is displayed instead.
If the option
not defined
is selected as
Device name
, all device types being
able to execute the command can be selected independently from the de-
vices in the device table.
Dosing device
Dosing device
[ 1 ] ... 4
(835, 836, 842, 857)
[ 1 ] ... 3
(855)
[ internal D0 ]
(Titrino)
external D1, external D2
(736, 751, 799)
Selection of the dosing device to be used for the dosing. All the dosing de-
vice connections (MSB) are displayed which are possible with the selected
device type.
Solution
24 characters, solution name, [ not defined ]
Input or selection of the solution defined in the solution table. Additionally,
not defined
can be selected. If intelligent exchange or dosing units are used
then a check will be made in the method sequence that the correct solution
is present in the connected dosing device and whether the type of dosing
device is correct. With non-intelligent exchange or dosing units the cylinder
volume is checked. At the start of the command the validity of the titer and
the expiry date of the selected solution will be checked as well as the GLP
test interval for the buret. If
not defined
is selected no tests will be carried
out.
Tandem dosing
on, [ off ]
(only Titrando)
If this option is enabled, the dosing is carried out without interruption with a
combination of two dosing devices so that the second dosing device is
dosing while the first one is being filled and vice versa (see STAT pH - Gen-
eral/Hardware).
Dosing device
1 ... [ 2 ] ... 4
(835, 836, 842, 857)
1 ... [ 2 ] ... 3
(855)
Selection of the second dosing device (exchange or dosing unit) to be
used for the dosing if the first dosing device is not available.
Solution
24 characters, solution name, [ not defined ]
(only Titrando, 855)
Input or selection of the solution defined in the solution table. Additionally,
not defined
can be selected. If intelligent exchange or dosing units are used
then a check will be made in the method sequence that the correct solution
is present in the connected dosing device and whether the type of dosing
device is correct. With non-intelligent exchange or dosing units the cylinder
volume is checked. If
not defined
is selected no tests will be carried out.