Metrohm tiamo 1.3 Manual User Manual
Page 463

Commands
437
Start drift
1 ... [ 20 ] ... 999 µg/min
The determination can only be started when the measured drift is below the
entered value. The Status in the
Run
window (Single determination or De-
termination series) then switches to
COND READY
.
Drift correction
automatic, manual, [ off ]
If this parameter is set to
automatic
or
manual
the drift correction ( =
Drift
value
x Titration time) caused by blank consumption during titration is de-
ducted from the total consumption. With
automatic
the drift value at the start
of the titration applies with
manual
the value defined under
Drift value
is ap-
plied.
Drift value
[ 0 ] ... 99.9 µL/min
Drift value for drift correction. This parameter is only shown for
Drift correc-
tion
=
manual
.
Delay after 'Cond ok'
[ 0 ] ... 999 s
Waiting period until the start of the titration during which the conditioning
condition must be fulfilled.
Sample addition time
[ 0 ] ... 99 s
Waiting time between stop of conditioning and start of titration during which
the sample can be added.
Only start titration by a start command from a SEND command
on, [ off ]
If this option is set the titration is only started when the
Event message
Start
titration
from a
SEND
command reaches this command.
Note
At the time of reception of the
SEND
command conditioning must already be
terminated (=
Condok
) otherwise the command to start the titration will be ig-
nored.
Show measured value during conditioning
on, [ off ]
If this option is set additionally to the drift and time the measured value is
shown in the Live display.