Subwindow Run
115
signed in a command then the window
Device assignment
appears for as-
signing the device.
•
Check solutions
Checks whether the solutions demanded in the device-specific commands
are present and available.
•
Check sensors
Checks whether the sensors demanded in the device-specific commands
are present and available.
Note
The Start test can also be triggered manually in the
READY
status with the menu
item
Tools, Run test
.
4 - Preconditioning (only for methods with conditioning)
In methods that contain commands with switched-on conditioning and in which
the option
Automatic conditioning
is switched on in the
START
command then pre-
conditioning will start for all these commands. Preconditioning does not belong to
the determination itself and does not increase any counter (
Sample number
,
Statis-
tics
).
•
Conditioning requirement not fulfilled
After the start of preconditioning the status changes to
COND BUSY
. In this
status conditioning can be started with
[Start]
even though the conditioning
requirement has not been fulfilled. Conditioning can also be canceled with
[Stop]
or interrupted with
[Hold]
and then continued with
[Cont]
. If an error
occurs during conditioning then neither the Error track nor the Exit track will
be carried out.
•
Conditioning requirement fulfilled
If the conditioning requirement has been fulfilled then the status changes to
COND READY
. In this status the determination can be started with
[Start]
.
Conditioning can also be canceled with
[Stop]
or interrupted with
[Hold]
and
then continued with
[Cont]
. The conditioning requirement will continue to be
monitored continuously until the determination has been started.
5 - Start method
At the start of the method the following checks and actions are triggered:
•
Check sample data
Checks that the entered Sample data corresponds to that defined in the
Method variables and is valid.
•
Assign determination ID
An unambiguous determination ID is assigned for the determination.
•
Reserve device (module)
The devices (or their functional units) used by the device-dependent com-
mands contained in the method are reserved for the duration of the deter-
mination, i.e. they cannot be used in other workplaces nor can they be op-
erated manually.
•
Increment sample number
The
Sample number
is increased by
+1
.
•
Increment statistics counter
If statistics has been switched on in both the
START
command and on the
tab
Single determinations
then the statistics counter is increased by
+1
. If a
new method is loaded or if the statistics counter agrees with the set counter
then the current statistical data will first be deleted and the statistics counter
reset to
0
.