Metrohm tiamo 1.3 Manual User Manual

Page 486

background image

460

Method

Rate evaluation

on, [ off ]

With this option a maximum of 9 time windows can be defined in which the

mean dosing rate can be determined by linear regression. The defined time
windows are shown in the window table and can be edited with the follow-
ing buttons:

Open the dialog window

Evaluation window #

(see below) to enter the pa-

rameters of a new time window.

Open the dialog window

Evaluation window #

(see below) to edit the pa-

rameters of the selected time window.

Delete the selected time window.

Evaluation window

Evaluation windows are defined time segments in which the dosing rate is deter-
mined by linear regression.

Lower limit

[ 0.0 ] ... 99999.9 s

Lower limit of evaluation window.

Upper limit

0.0 ... [ 999999.9 ] s

Upper limit of evaluation window.

Fix end point evaluation

on, [ off ]

For a fixed quantity (measured value, volume or time) for the fix end point
the associated values are interpolated from the list of measuring points.

Fixed quantity

[ Measured value ], Time, Volume

Selection of the fixed quantity to which the associated values for the other
quantities are to be interpolated from the list of measured points.

Fix EP# at pH

-20.000 ... 20.000, [ off ]

Fix end point 1 ... 9 for

Fixed quantity

=

Measured value

.

Fix EP# at

0.0 ... 999999.9 s, [ off ]

Fix end point 1 ... 9 for

Fixed quantity

=

Time

.

Fix EP# at

0.00000 ... 9999.99 mL, [ off ]

Fix end point 1 ... 9 for

Fixed quantity

=

Volume

.