Commands

479

For the minimum measured value the associated volume, time and tem-
perature are interpolated from the list of measuring points (see Titration

commands - Evaluation - Minimum/maximum evaluation).

Threshold value

1.0 ... [ 25.0 ] ... 2000.0 mV/mL

The evaluation begins as soon as the slope of the curve exceeds the set
threshold value.

Maximum evaluation

on, [ off ]

For the maximum measured value the associated volume, time and tem-
perature are interpolated from the list of measuring points (see Titration
commands - Evaluation - Minimum/maximum evaluation
).