7 calibration commands – Metrohm tiamo 1.3 Manual User Manual

Page 559

background image

Commands

533

mol/L, mmol/L, µmol/L, [ ppm ], g/L, mg/L, µg/L, mg/mL, µg/mL, %, mEq/L

Concentration unit of addition solution.

Control parameters

Dosing rate

slow, medium, [ fast ]

Speed at which the volume increments are to be added.

Delta U

1 ... [ 10 ] ... 999 mV

Potential difference that is to be achieved by a standard addition.

Measuring parameters

Screenshot and parameters: see STDADD man - Measuring parameters

5.6.7 Calibration

commands

Commands for calibration of pH- and ion selective sensors.

General

The commands

CAL LOOP

and

CAL MEAS

are available for calibrating electrodes.

The number and type of buffers or standards respectively is defined in the loop
command

CAL LOOP

whereas the measurement is done with the command

CAL

MEAS

. The validity of the measurement is checked after each pass of the loop. The

calibration data for the electrode are determined at the end.

The basic structure of a calibration method looks as follows: