Metrohm tiamo 1.3 Manual User Manual
Page 262

236
Database
Calibration curve for CAL LOOP pH
In the dialog window
Calibration curve
a tab marked with the command name and
showing the calibration curve and calibration data will be shown for each
CAL
LOOP pH
command. The command type is shown above the calibration curve. The
curve display shows the measured values and the curve calculated from these
measured values. Below it are listed the individual measured values and the re-
sults for
Slope
,
E(0)
and
Variance
.
Calibration curve for CAL LOOP Conc
In the dialog window
Calibration curve
a tab marked with the command name and
showing the calibration curve and calibration data will be shown for each
CAL
LOOP Conc
command. The command type is shown above the calibration curve.
The curve display shows the measured values and the curve calculated from
these measured values. Below it are listed the individual measured values and the
results for
Slope
,
E(0)
,
c(blank)
and
Variance
.
Calibration curve for STDADD
In the dialog window
Calibration curve
a tab marked with the command name and
showing the standard addition curve and its evaluation data will be shown for
each STDADD command. The command type is shown above the calibration
curve. The curve display shows the measured values and the curve calculated
from these measured values. Below it are listed the individual measured values
and the results for
Slope
,
E(0)
,
ion concentration
and
Variance
.
Note
Calibration curves cannot be printed out directly from the dialog window
Calibra-
tion curve
. Instead you should use a report with a Report template in which the
fixed report
Calibration curve
is contained.