792

Configuration

life of the sensor has expired will be saved automatically with the
determination.

Stop determination

The running determination will be terminated automatically. The following
message must be confirmed with

[OK]

.

Calibration data

Note
This tab only appears for pH and ISE electrodes.

Slope

pH electrode:

-999.9 ... 999.9 %, [ 100.0 ]

ISE electrode:

-999.9 ... 999.9 mV, [ 59.2/valency ]

Slope of the pH or ISE electrode. This value can be determined automati-

cally with a calibration from the linearized calibration curve or entered
manually. The slope of a sensor that is used by the commands

DET

,

MET

,

SET

,

KFT

,

MEAS

and

CAL LOOP

is available as a variable

'Command name.SLO'

for calculations.

pH(0)

-20.000 ... [ 7.000 ] ... 20.000

The electrode zero point

pH(0)

is only shown for pH electrodes. After the

slope the

pH(0)

is the second characteristic quantity of the calibration func-

tion U = f(pH).

pH(0)

is the x-axis intercept of the calibration curve, i.e. it

corresponds to the pH value at 0 mV. The electrode zero point

pH(0)

of a

pH electrode that is used by the commands

DET

,

MET

,

SET

,

KFT

,

MEAS

and

CAL LOOP

is available as a variable

'Command name.ENP'

for calculations.

E(0)

-2000.0 ... [ 0.0 ] ... 2000.0 mV

The electrode zero point

E(0)

is only shown for ISE electrodes.

E(0)

is the

second characteristic quantity of the calibration function U = f(log c).

E(0)

is the y-axis intercept of the calibration curve and corresponds to the poten-

tial at log c = 0. The electrode zero point

E(0)

of an ISE electrode that is

used by the commands

DET

,

MET

,

SET

,

KFT

,

MEAS

and

CAL LOOP

is available

as a variable

'Command name.ENP'

for calculations.