Subwindow Devices
707
Data storage under legal control
on, [ off ]
This option must be enabled if the balance has its own data memory (only
for Sartorius balances).
Remarks
1000 characters, [ empty ]
Remarks about the device.
This button is only shown with a balance or barcode reader. In order to de-
lete the device from the device list the device must be disconnected.
Load new program version
If when a device is connected up it is discovered that this has an old program ver-
sion that is not supported by tiamo then it must be updated. This is done by se-
lecting the device in the Device table and using
Edit, Properties
to open the Proper-
ties window for the device. On the tab
General
the button
[Update]
is then active.
Opens the dialog window
Load new program version
.
Old version
Shows the old device program version.
New version
Shows the new device program version to be loaded.
Loads new program version.
Note
Make sure that the device is not manipulated or switched off during the loading
process and follow the instructions shown.
RS 232
COM Port
COM1, COM2, ..., [ first free COM port ]
Selection of the serial interface on the PC to which the device is connected.
Baud rate
1200, 2400, 4800, [ 9600 ]
Transmission speed. Additionally, this baud rate must be set on the device
itself.