Metrohm tiamo 1.3 Manual User Manual
Page 378
352
Method
Device type
Device types, [ Titrando ]
Display or selection of the device type. If a device is selected under
Device
name
, this parameter is not editable any more. The device type belonging to
the selected device is displayed instead.
If the option
not defined
is selected as
Device name
, all device types being
able to execute the command can be selected independently from the de-
vices in the device table.
Dosing device
Dosing device
[ 1 ] ... 4
(808, 809, 835, 836, 857)
[ 1 ] ... 3
(855)
[ 1 ]
(888)
[ internal D0 ]
(Titrino)
external D1, external D2
(only 736, 751, 758, 799)
Selection of the number of the dosing device (dosing or exchange unit) to
be used for the dosing.
Solution
24 characters, solution names, [ not defined ]
Input or selection of the titrant or solution defined in the solution table. Addi-
tionally,
not defined
can be selected. If intelligent exchange or dosing units
are used then a check will be made in the method sequence that the cor-
rect titrant is present in the connected dosing device and whether the type
of dosing device is correct. With non-intelligent exchange or dosing units
the cylinder volume is checked. At the start of the command the validity of
the titer and the expiry date of the selected titrant will be checked as well as
the GLP test interval for the buret. If
not defined
is selected no tests will be
carried out.
Sensor
Measuring input
[ 1 ] ... 2
(808, 809, 835, 836, 857)
[ 1 ]
(855, 888)
[ 1 ] ... 2, diff.
(Titrino)
Selection of the measuring input to which the sensor is connected.
Sensor
Sensor name, [ Metal electrode ]
Selection of a sensor of type
pH electrode, Metal electrode
or
ISE electrode
from the list of sensors defined in the sensor table. The calibration data for
the ISE and pH sensors will be adopted for the determination.
Temperature measurement
Type of temperature measuring (only Titrando):
continuous
A temperature sensor must be connected. The temperature is measured
continuously.
[ automatic ]
If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise the manually entered
Temperature
(tab
Titra-
tion parameters
) will used.
off
The temperature will not be measured. The
Temperature
(see Titration pa-
rameters) which has been entered manually will be used.
Stirrer
Stirrer
[ 1 ] ... 4, off
(only Titrando)