beautypg.com

Metrohm tiamo 1.3 Manual User Manual

Page 478

background image

452

Method

Last dosing

The stop time starts after the last increment has been dosed, i. e. the stop

time is reset to zero again after each dosing step.

Stop rate

0.01 ... 166.00 mL/min, [ off ]

(Titrando)

0.01 ... 150.00 mL/min, [ off ]

(Titrino)

Stops when the current titration rate is lower than the rate defined here. This
parameter is checked 10 s after the start at the earliest. The maximum stop

rate depends on the cylinder volume of the exchange unit or dosing unit
used.

Filling rate

0.01 ... 166.00 mL/min, [ maximum ]

(Titrando)

0.01 ... 150.00 mL/min, [ maximum ]

(Titrino)

Speed with which the buret is to be refilled after the titration. The maximum
filling rate depends on the cylinder volume of the exchange unit or dosing
unit used.

Monitoring

The following monitoring processes can be defined and switched on here: