Metrohm tiamo 1.3 Manual User Manual
Page 808
782
Configuration
Tubing parameters
Defines the length and diameter of the tubing connected to the dosing unit. The
port occupancy can also be altered. These parameters are important for carrying
out the dosing unit commands
PREP
and
EMPTY
correctly, as the volumes of the
tubing connections have to be taken into account.
Note
Default values have already been entered for the tubing parameters; these cor-
respond to the dimensions of the standard tubing supplied. If you do not make
any alterations to the tubing connections then you do not need to alter the tubing
parameters. For information on length und diameter of other tubings see
Acces-
sories on www.metrohm.com/products.html.
Dosing port 1
Port
[ Port 1 ], Port 2, Port 3, Port 4
Port to be used as Dosing port 1.
Length
0.0 ... [ 40.0 ] ... 999.9 cm
Length of tubing attached to Dosing port 1.
Diameter
0.0 ... [ 2.0 ] ... 9.9 mm
Diameter of tubing attached to Dosing port 1.
Dosing port 2
Port
Port 1, Port 2, [ Port 3 ], Port 4
Port to be used as Dosing port 2.
Length
[ 0.0 ] ... 999.9 cm
Length of tubing attached to Dosing port 2.
Diameter
0.0...[ 2.0 ]...9.9 mm
Diameter of tubing attached to Dosing port 2.
Fill port
Port
Port 1, [ Port 2 ], Port 3, Port 4
Port to be used as Fill port for aspirating the solution.
Length
0.0...[ 25.0 ]...999.9
Length of tubing attached to Fill port.
Diameter
0.0...[ 2.0 ]...9.9 mm
Diameter of tubing attached to Fill port.
Special port
Port
Port 1, Port 2, Port 3, [ Port 4 ]
Port to be used as Special port.
Length
[ 0.0 ] ... 999.9 cm
Length of tubing attached to Special port.
Diameter
0.0...[ 2.0 ]...9.9 mm
Diameter of tubing attached to Special port.