beautypg.com

Metrohm tiamo 1.3 Manual User Manual

Page 394

background image

368

Method

Minimum difference between the slope before and after the break point.
The smaller the difference, the more break points will be found.

Smoothing factor

2 ... [ 5 ] ... 20

The higher the smoothing factor, the fewer break points will be found.

Window

Measured value, Volume, Time, [ off ]

A range (window) can be defined on the measured value axis, on the vol-
ume axis or on the time axis. The break point evaluation will only be carried

out in the defined window. Only the first break point in the defined window
will be recognized.

Lower limit

[ -2000.0 ] ... 2000.0 mV

(for

Window

=

Measured value

)

[ 0.00000 ] ... 9999.99 mL

(for

Window

=

Volume

)

[ 0 ] ... 999999 s

(for

Window

=

Time

)

Value for the lower limit of the window.

Upper limit

-2000.0 ... [ 2000.0 ] mV

(for

Window

=

Measured value

)

0.00000 ... [ 9999.99 ] mL

(for

Window

=

Volume

)

0 ... [ 999999 ] s

(for

Window

=

Time)

Value for the upper limit of the window.

Gran evaluation

on, [ off ]

(only Titrando)

With the Gran evaluation titration curves are evaluated according to the
Gran plot procedure (see Titration commands - Evaluation - Gran evalua-
tion
).

Initial volume

0.01 ... [ 50.00 ] ... 9999.99 mL

Volume which must be present in the measuring vessel before the com-
mand is started.

Lower limit

[ -2000.0 ] ... 2000.0 mV

Value for the lower limit of the Gran evaluation range.

Upper limit

-2000.0 ... [ 2000.0 ] mV

Value for the upper limit of the Gran evaluation range.

Additional measured values

Screenshot and parameters: see DET pH - Additional measured values