Commands
401
Conditioning
Command name
25 characters, [ SET pH # ]
Name of the command.
Conditioning
[ on ], off
Switching conditioning on/off. If conditioning is deactivated all following pa-
rameters are not visible.
Start drift
1 ... [ 20 ] ... 999 µL/min
(only Titrando)
The determination can only be started when the measured drift is below the
entered value. The Status in the Run window (Single determination or De-
termination series) then switches to
COND READY
.
Drift correction
automatic, manual, [ off ]
If this parameter is set to
automatic
or
manual
the drift correction ( =
Drift
value
x Titration time) caused by blank consumption during titration is de-
ducted from the total consumption. With
automatic
the drift value at the start
of the titration applies with
manual
the value defined under
Drift value
is ap-
plied
Drift value
[ 0 ] ... 99.9 µL/min
Drift value for drift correction. This parameter is only shown for
Drift correc-
tion
=
manual
.
Stop volume
0.000000 ... [ 20.00 ] ... 9999.99 mL, [ off ]
(only Titrando)
Stops when the given volume has been added after the start of the titration.
This value should be adapted the volume of the sample used to prevent
overflow.
Stop time
0 ... 999999 s, [ off ]
(only Titrando)