730 sample changer – Metrohm tiamo 1.3 Manual User Manual
Page 767
Subwindow Devices
741
If this option is switched on then the time interval for the GLP test will be
monitored.
Note
This option can only be switched on when a date has been entered under
GLP
test date
.
GLP test interval
0 ... [ 999 ] days
Time interval to next GLP test. If an interval is entered here then the date for
the
Next GLP test
will be adapted automatically.
Next GLP test
Date selection, [GLP test date + 999 days ]
Date on which the next GLP test is to be carried out. This date can be se-
lected after pressing
in the dialog window
Select date
(see General
program functions - Edit - Select date). After the date has been entered the
value for the
GLP test interval
will be adapted automatically.
Action
If the time interval for the GLP test has expired then one of the following actions
will be triggered at both the Start test and the command sequence:
Record message
The message that the GLP test interval has expired will be saved automati-
cally in the determination.
Display message
A message appears and you can then choose whether to continue with the
run or to terminate it. If the run is continued then the message that the GLP
test interval has expired will be saved automatically in the determination.
Stop determination
The running determination will be automatically terminated. The following
message must be confirmed with
[OK]
.
730 Sample Changer
The parameters for the 730 Sample Changer are set on the following tabs:
•
General
General device information such as device name, device type, serial num-
ber, etc.
•
Towers
Properties of Tower 1 and Tower 2 (if present).
•
Rack
Information about the attached rack.
•
Dosing devices
Properties of the connection and the dosing devices connected to it.
•
RS232
Selection of the serial interface to which the device is connected.
•
GLP
Information about GLP test and GLP monitoring.