Page 778

752
Configuration
Shift position
[ 0 ] ... 100 mm
Shift position for Lift 1. Each time that the rack rotates the lift will move to
this position if it is located at a lower lift position. If the lift is located at a
higher lift position than that defined here then the rotation will take place at
the current lift position. This means that the shift position must be selected
so that a safe rotation over the whole rack is possible at any time.
Special position
[ 0 ] ... 100 mm
Special position for Lift 1. This additional definable position can be used
e.g. during pipetting so that the tip is just immersed in the sample solution.
Special beakers
Special beaker
Shows the number of the special beaker for the selected rack.
Rack position
Shows the number of the rack position for the special beaker.
Dosing device
Shows the number and type of dosing devices connected to the 730 Sample
Changer or 774 Oven Sample Processor. A maximum of 12 type
685 Dosimat
or
700 Dosino
dosing devices can be connected.
Note
Connecting up the dosing devices is described in the Instructions for Use for the
730 Sample Changer and 774 Oven Sample Processor. If no dosing device is
connected then the card will remain empty.