Sample data – Metrohm tiamo 1.3 Manual User Manual
Page 149
Subwindow Run
123
The first field after
Statistics
shows the number of determinations that have
already been carried out for the statistics (actual counter). This field cannot
be edited, but in the status
READY
the context-sensitive menu item
Delete
statistics
can be used to reset it to
0
, which at the same time deletes the
statistics data that has already been calculated.
The second field after
Statistics
shows the number of determinations to be
carried out for the statistics (set counter). It contains the value for the
Num-
ber of single determinations
defined in the
START
command as standard.
This field can only be edited in the status
READY
.
Modify remark
With the context-sensitive menu item
Modify remarks
the dialog window
Live modifi-
cations
opens, in which the remark can be edited live during a running determina-
tion.
Remark
100 characters
Freely definable comments about the determination that are saved together
with the determination. Instead of a manual entry it is also possible to select
from prepared Text templates.
Modify autostart counter
With the context-sensitive menu item
Modify autostart counter
the dialog
window
Live modifications
opens, in which the autostart counter can be al-
tered live during a running determination.
Autostart
1 ... 999, [ Sample table ]
Number of automatic, internal starts of determinations within a sample se-
ries.
The first non-editable field after
Autostart
shows the number of determina-
tions that have already been carried out (actual counter).
The second editable field after
Autostart
shows the total number of determi-
nations to be carried out per series (set counter).
Sample data
Under
Sample data
the current working sample table is shown that contains the
sample data for the determination series in tabular form. Each line corresponds to
a single determination.
Working sample table
The table contains the sample data lines already saved in the sample table as well
as a line for entering new data which contains an asterisk instead of the line num-
ber. The table cannot be directly edited or sorted.
For the meaning of the columns please refer to Edit sample data.
The lines in the working sample data table can have different background colors: