Calibrating with automatic solution changing – Metrohm tiamo 1.3 Manual User Manual
Page 562
536
Method
Calibrating with automatic solution changing
The basic structure of a calibration method with automatic solution exchange on a
Sample Processor looks as follows:
Defining the number of
buffers/standards and type
of solution exchange.
Move to the next
buffer/standard.
The request for changing the buffer/standard must be deactivated in the com-
mand
CAL LOOP pH
or
CAL LOOP Conc
respectively. No request for changing the
buffer/standard is displayed during the method run.
For the automatic move to the next calibration solution on the Sample Processor a
command
MOVE
must be inserted between the loop command
CAL LOOP
and the
measuring command
CAL MEAS
. The positions of the calibration solutions on the
sample rack are defined in the command
MOVE
. There are three possibilities:
•
Defining rack positions for calibration solutions in the method
A
Rack position
###
for the first calibration solution has to be defined in the
command
MOVE
. The other calibration solutions must be placed on the di-
rectly following rack positions. The following settings have to be done in the
command
MOVE
under the parameter
Target
:
Move
:
Rack position
Number
:
= ### – 1 + 'Command name.LCO'
(entry with Formula editor, i.e.
=
19 + 'Calibration loop pH.LCO'
for the above sample method with the first