Commands

329

Evaluation

Equivalence points (EPs) are determined atuomatically in a similar manner to that

described in the Tubbs method (C.F. Tubbs; Anal. Chem 26 (1954) 1670–1671,
quoted in Ullman 5 (1980) 659). The volume of the equivalence point (V

E

) is shifted

from the point of inflection towards the smaller circle of curvature for real asym-
metric titration curves.


The figure shows that the evaluation also requires measuring points after the
equivalence point too.
For the recognition of the EPs found the set

EP criterion

is compared with the

ERC

(

E

quivalence point

R

ecognition

C

riterion) found. The ERC is the first derivative of

the titration curve combined with a mathematical function which is more sensitive
for flat jumps than for steeper ones. EPs whose ERC is smaller than the set EP cri-
terion will not be recognized.

Commands

Depending on the measured value one of the following four

DET

commands can

be chosen:

DET pH
Potentiometric pH measurement with pH electrodes (measured quantity
pH).

DET U
Potentiometric voltage measurement with metal electrodes (measured
quantity voltage U).

DET Ipol
Voltametric measurement with selectable polarization current (measured

quantity voltage U).