Metrohm tiamo 1.3 Manual User Manual
Page 402
376
Method
der
Sensor
is set to
automatic
or
continuous
then the temperature will be
measured continuously.
Stop conditions
Conditions for stopping the titration. The stop conditions are checked at the same
time and the condition which is fulfilled first stops the titration.
Command name
25 characters, [ DET Upol # ]
Name of the command.
Stop volume
0.00000 ... [ 100.000 ] ... 9999.99 mL, off
(Titrando)
0.000 ... [ 100.000 ] ... 9999.99 mL, off
(Titrino)
Stops when the given volume (including start conditions) has been added
after the start of the titration. The stop volume should be adapted to suit the
sample weight or the titration vessel size.
Stop measured value
-200.0 ... 200.0 µA, [ off ]
Stops when the preset value for a measuring point has been exceeded or
undercut since the start of the titration.
Stop EP
1 ... [ 9 ], off
Stops when the given number of EPs has been found..
Volume after EP
0.01000 ... 9999.99 mL, [ off ]
(only Titrando)
When the number of equivalence points defined under
Stop EP
has been
found this volume will be added. In this way the curve can be continued af-
ter the equivalence point has been reached.
Stop time
0 ... 999999 s, [ off ]
(only Titrando)
Stops when the preset time (including start conditions) has elapsed since
the start of the titration.
Filling rate
0.01 ... 166.00 mL/min, [ maximum ]
(Titrando)
0.01 ... 150.00 mL/min, [ maximum ]
(Titrino)
Speed with which the buret is to be refilled after the titration. The maximum
filling rate depends on the cylinder volume of the exchange unit or dosing
unit used.