Metrohm tiamo 1.3 Manual User Manual
Page 566
540
Method
Command name
25 characters, [ CAL MEAS pH # ]
Name of the command.
Device
Device name
Device name, [ not defined ]
Selection of a device from the device table. Only those devices being able
to execute the command are selectable. Additionally,
not defined
can be se-
lected. In this case, the device name must be assigned by the user at the
start of the method.
Device type
Device types, [ Titrando ]
Display or selection of the device type. If a device is selected under
Device
name
, this parameter is not editable any more. The device type belonging to
the selected device is displayed instead.
If the option
not defined
is selected as
Device name
, all device types being
able to execute the command can be selected independently from the de-
vices in the device table.
Sensor
Measuring input
[ 1 ] ... 2
(Titrando, Titrino)
[ 1 ]
(855)
Selection of the measuring input to which the sensor is connected.
Sensor
Sensor name, [ pH electrode ]
Selection of a sensor of type
pH electrode
from the list of sensors defined in
the sensor table.
Temperature measurement
Type of temperature measurement (Titrando only):
continuous
A temperature sensor must be connected. The temperature is then meas-
ured continuously.
[ automatic ]
If a temperature sensor is connected, the temperature will be measured
continuously. Otherwise the temperature manually entered under
Tempera-
ture
(see Measuring parameters) will be used.
off