Method

Command name

25 characters, [ CAL MEAS pH # ]

Name of the command.

Device

Device name

Device name, [ not defined ]

Selection of a device from the device table. Only those devices being able
to execute the command are selectable. Additionally,

not defined

can be se-

lected. In this case, the device name must be assigned by the user at the
start of the method.

Device type

Device types, [ Titrando ]

Display or selection of the device type. If a device is selected under

Device

name

, this parameter is not editable any more. The device type belonging to

the selected device is displayed instead.
If the option

not defined

is selected as

Device name

, all device types being

able to execute the command can be selected independently from the de-
vices in the device table.

Sensor

Measuring input

[ 1 ] ... 2

(Titrando, Titrino)

[ 1 ]

(855)

Selection of the measuring input to which the sensor is connected.

Sensor

Sensor name, [ pH electrode ]

Selection of a sensor of type

pH electrode

from the list of sensors defined in

the sensor table.

Temperature measurement

Type of temperature measurement (Titrando only):

continuous

A temperature sensor must be connected. The temperature is then meas-
ured continuously.

[ automatic ]

If a temperature sensor is connected, the temperature will be measured
continuously. Otherwise the temperature manually entered under

Tempera-

ture

(see Measuring parameters) will be used.

off