Formula editor
23
Identifier
Description
Commands
.ENP
Electrode zero point of the sensor used for the command (di-
mensionless for pH sensor or in mV for ISE sensor) or the elec-
trode zero point calculated from the calibration (for CAL LOOP)
DET pH, DET U,
MET pH, MET U,
SET pH, SET U,
STAT, MEAS pH,
MEAS U, MEAS T,
MEAS Conc,
STDADD, CAL
LOOP, DOS
.EP{#}.DME
Difference in measured values for end point # (1...9)
MET
.EP.DVT
Drift for end point in µg/min
KFC
.EP{#}.ERC
ERC for the end point # (1...9)
DET
.EP{#}.MEA
Measured value for the end point # (1...9) in the units of the
measured value
DET, MET, SET,
KFT, KFC
.EP{#}.MEP
Number of end points in the window # (1...9); 1 = 1 end
point, 2 = 2 or more end points, 3 = EP corrected with
auto drift, 4 = EP corrected with manual drift
DET, MET, SET,
KFT
.EP.QTY
Measured value (water) for end point in µg
KFC
.EP{#}.TEM
Temperature for end point # (1...9) in °C
DET, MET. SET,
KFT
.EP{#}.TIM
Time for reaching end point # (1...9) in s
DET, MET, SET,
KFT, KFC
.EP{#}.VOL
Volume for end point # (1...9) in mL
DET, MET, SET,
KFT
.ETE
Final temperature (temperature after processing the com-
mand) in °C
DET, MET, SET,
KFT, KFC, STAT,
MEAS (without
T/Flow), CAL
MEAS, STDADD,
DOS
.EVT
Final volume (total volume added at the end of the command)
in mL
DET, MET, SET,
KFT, STAT, DOS
.FIN
Command status; 1 = command has been finished at least
once, 0, invalid (variable not present) = command has
never been finished
All except START
and END
.FP{#}.DME
Difference in measured values for fixed end point # (1...9)
MET
.FP{#}.DVT
Drift for fixed end point # (1...9) in µg/min
KFC
.FP{#}.ERC
ERC for fixed end point # (1...9)
DET
.FP{#}.MEA
Measured value for fixed end point # (1...9) in mV
DET, MET, SET,
KFT, STAT
.FP{#}.MEA
Measured value for fixed end point # (1...9) in the units of the
measured value
KFC, MEAS (with-
out T/Flow)
.FP{#}.QTY
Measured value (water) for fixed end point # (1...9) in µg
KFC
.FP{#}.TEM
Temperature for fixed end point # (1...9) in °C
DET, MET, SET,
KFT, STAT, MEAS
(without T/Flow)